Previous Article | Next Article ![]()
Antimicrobial Agents and Chemotherapy, March 2000, p. 771-772, Vol. 44, No. 3
Department of
Microbiology,1 Department of Animal
Science,2 Department of Chemical
Engineering,3 and Department of
Statistics,4 Brigham Young University, Provo,
Utah
Received 2 July 1999/Returned for modification 21 September
1999/Accepted 29 November 1999
Escherichia coli biofilms on two polyethylene disks
were implanted subcutaneously into rabbits receiving systemic
gentamicin. Ultrasound was applied for 24 h to one disk. Both
disks were removed, and viable bacteria were counted. Pulsed ultrasound
significantly reduced bacterial viability below that of nontreated
biofilms without damage to the skin.
Successful treatments of infections
on implanted medical devices are rare, even with extensive antibiotic
therapy (1, 3, 4). Recently our laboratory reported that the
combination of systemic gentamicin and application of low-frequency
continuous ultrasound to a simulated implant infection in a rabbit
model significantly reduced the concentration of viable bacteria on the
implant (6). However, the power density required to reduce the concentration (300 mW/cm2) also caused some damage to
the skin. Herein, we report that using pulsed ultrasound and gentamicin
significantly reduces biofilm viability without causing any apparent
damage to the skin.
Seven New Zealand White female rabbits (2.0 to 2.5 kg) were treated in
accordance with the regulations of the Institutional Animal Care and
Use Committee of Brigham Young University and the U.S. Department of
Agriculture. Procedures for rabbit maintenance, measurement of
subcutaneous heating, implantation surgery, and evaluation of biofilm
viability were reported previously (6) and are described
briefly below.
Escherichia coli strain ATCC 10798 was grown in biofilms on
sterile circular polyethylene disks with two sewing tabs. This strain
produced a thick biofilm in 24 h of growth (2). After the rabbit was anesthetized and its back was denuded of fur, these disks were implanted subcutaneously and sewn into place, one on each
side of the spine. The rabbit received gentamicin (8 mg/kg) by
subcutaneous injection following surgery and every 24 h
thereafter. Blood was sampled from the ear just prior to each
gentamicin injection and prior to euthanization.
A 49-g ultrasound transducer (Tonpilz resonator, model 6147E; EDO
Acoustics, Salt Lake City, Utah) operating at 28.48 kHz was fixed over
one implant 24 h postsurgery. Unlike previous experiments with
continuous ultrasound, the ultrasound was delivered in a pulse of 100 cycles with a 1:3 or 1:6 duty cycle wherein the temporal average
intensity was 100 mW/cm2. For example, ultrasound at 300 mW/cm2 was pulsed in a 1:3 duty cycle in which the
ultrasound was on for 3.51 ms (100 cycles) and off for 7.02 ms (average
intensity of 100 mW/cm2). Likewise, ultrasound at 600 mW/cm2 was pulsed in a 1:6 duty cycle: on for 3.51 ms and
off for 17.56 ms. This experimental design could show whether enhanced
killing is a function of average intensity or peak intensity during the pulse.
Following 24 h of pulsed ultrasound, rabbits were euthanized, both
disks were recovered, and the viable bacterial concentration was
determined. Viable counts were analyzed using a one-tailed t
test of the mean difference in paired observations of treated and
nontreated sides. Skin and tissue from the implant site and heart,
liver, spleen, and kidney samples were taken for histopathology as
described previously (5, 6).
In an experiment without implantation of disks, pulsed ultrasound at
duty cycles of 1:3 and 1:6 raised the subcutaneous skin temperature to
37.7 and 38.2°C, respectively, and no skin damage was observed.
Our previously reported experiments showed that gentamicin combined
with continuous ultrasound at 100 mW/cm2 did not reduce
bacterial viability, whereas continuous ultrasound at 300 mW/cm2 reduced viability but caused skin damage (Fig.
1). Data from the 300-mW/cm2
pulsed-ultrasound (1:3 duty cycle) experiments also showed a significant decrease in viability (P = 0.017; n = 3), the average counts being reduced from 2.94 to 0.99 log10 CFU/cm2. More importantly, the skin
showed no discoloration or damage.
0066-4804/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Pulsed Ultrasound Enhances the Killing of
Escherichia coli Biofilms by Aminoglycoside Antibiotics
In Vivo
![]()
ABSTRACT
Top
Abstract
Text
References
![]()
TEXT
Top
Abstract
Text
References

View larger version (31K):
[in a new window]
FIG. 1.
Comparison of log10 numbers of viable CFU
from biofilm-coated polyethylene disks. The second bar from the right
represents data from experiments in which ultrasound (US) at 300 mW/cm2 was pulsed in a 1:3 duty cycle, producing a
time-averaged intensity of 100 mW/cm2. Likewise, the far
right bar represents experiments in which ultrasound at 600 mW/cm2 was pulsed in a 1:6 duty cycle, also producing a
time-averaged intensity of 100 mW/cm2. The asterisks
indicate sets of data that are significantly different (P < 0.05) from data for implants treated with gentamicin only.
Data from the 600-mW/cm2 pulsed-ultrasound (1:6 duty cycle) experiments also showed a significant viability reduction (P = 0.016; n = 3) in which average counts were reduced from 2.93 to 1.69 log10 CFU/cm2. Again, no skin damage was apparent.
Histopathology indicated necrosis, edema, a fibrous capsule, and inflammation in the skin and skeletal muscle tissues around both treated and nontreated implant sites. The histology of the skin receiving pulsed ultrasound was not unlike that of control skin or skin receiving continuous ultrasound at 100 mW/cm2. None of the blood samples taken before, during, or after ultrasound treatment showed the presence of E. coli bacteremia. The tissue around the implant site had some evidence of bacteria in both the ultrasound-treated and nontreated samples, but none of the other organs and tissues sampled showed any colonization by E. coli.
Comparison of all of the data from the pulsed-ultrasound and previous experiments shows that ultrasound enhances the action of gentamicin in killing E. coli when the pulse (or continuous) intensity is at least 300 mW/cm2. As Fig. 1 shows, viability is not reduced by continuous ultrasound at 100 mW/cm2 but it is reduced by continuous ultrasound at 300 mW/cm2 and by pulsed ultrasound at 300 or 600 mW/cm2. Enhancement does not appear to correlate with the temporal average intensity, since pulsed ultrasound at an average of 100 mW/cm2 enhanced killing while continuous ultrasound at the same average intensity did not. This suggests that the bacteria respond to the maximum or peak ultrasound intensity and not to the average intensity or the total amount of energy delivered.
Conversely, skin damage appears to correlate with average ultrasonic power and not maximum power during a pulse. An average intensity of 100 mW/cm2 produced no skin damage, whether pulsed or continuous, but an average intensity of 300 mW/cm2 produced significant damage. These results show that there is a window of operation in which one can use pulsed ultrasound of sufficiently high intensity (during the pulse) to enhance the action of the antibiotic while still maintaining the average intensity at a safe level that does not injure skin.
A final important observation about these higher-intensity ultrasonic pulses is that they do not appear to cause spread of bacteria from the infection site to other tissues. No viable E. coli cells were plated from blood, nor were any observed in histological sections of the internal organs.
Although we cannot report complete elimination of the infection, we are hopeful that higher pulse intensities and longer treatment times may lead to noninvasive treatment of implant infections.
| |
ACKNOWLEDGMENTS |
|---|
This research was funded by NIH grant HL59923.
The transducers and technical expertise were donated by Gordon Snow of EDO Acoustics. We thank David O. Draper of BYU for assistance in measuring the subcutaneous temperature.
| |
FOOTNOTES |
|---|
* Corresponding author. Mailing address: 350M Clyde Building, Brigham Young University, Provo UT 84602. Phone: (801) 378-2589. Fax: (801) 379-7799. E-mail: pitt{at}byu.edu.
| |
REFERENCES |
|---|
|
|
|---|
| 1. | Darouiche, R. O., A. Dhir, A. J. Miller, G. C. Landon, I. I. Raad, and D. M. Musher. 1994. Vancomycin penetration into biofilm covering infected prostheses and effect on bacteria. J. Infect. Dis. 170:720-723[Medline]. |
| 2. | Johnson, L., R. V. Peterson, and W. G. Pitt. 1998. Treatment of bacterial biofilms on polymeric implants using antibiotics and ultrasound. J. Biomater. Sci. Polym. Ed. 9:1177-1185[Medline]. |
| 3. | Khoury, A. E., K. Lam, B. Ellis, and J. W. Costerton. 1992. Prevention and control of bacterial infections associated with medical devices. Am. Soc. Artif. Intern. Organs J. 38:M174-M178. |
| 4. | Potera, C. 1996. Biofilms invade microbiology. Science 273:1795-1797[CrossRef][Medline]. |
| 5. | Rediske, A. M. 1999. Ultrasonic enhancement of antibiotic action on Escherichia coli biofilms: an in vivo model. M.S. thesis. Brigham Young University, Provo, Utah. |
| 6. |
Rediske, A. M.,
B. L. Roeder,
M. K. Brown,
J. L. Nelson,
R. L. Robison,
D. O. Draper,
G. B. Schaalje,
R. A. Robison, and W. G. Pitt.
1999.
Ultrasonic enhancement of antibiotic action on Escherichia coli biofilms: an in vivo model.
Antimicrob. Agents Chemother.
43:1211-1214 |
This article has been cited by other articles:
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»