Table 4.

Intralaboratory variation in IC50 for laboratory strain AD169 using the ACTG consensus assay

LaboratoryGCVPFA
na IC50 (μM)Mean (μM)CVbnIC50 (μM)Mean (μM)CV
A34.405.7660.5369.6683.8015.5
3.1786.55
9.7395.19
B31.8627.9156.83184.22338.6781.7
78.49657.39
3.44173.86
C43.673.8520.5243.9057.9534.3
4.3772.00
2.82
4.56
F23.182.7124.5270.8155.1540.2
2.2439.48
I31.403.1448.5376.7498.4219.3
4.24111.87
2.68106.63
K45.635.3928.5371.2885.8731.9
6.0768.87
3.18117.45
6.69
  • a Number of assays performed in each laboratory.

  • b CV determined as (standard deviation/mean) × 100%.